Direct Measurement of the Ciliary Sulcus Diameter Using Optical Coherence Tomography—Inter-Rater Variability

Acceso Abierto

Abstract:

The determination of sulcus-to-sulcus measurements has been challenging due to the limitations of current approaches. Ultrasound methods are highly operator-dependent and require extensive training, while traditional optical devices cannot visualize structures posterior to the iris. However, modern optical anterior segment coherence tomography (AS-OCT) devices are changing this paradigm by identifying some anatomical landmarks posterior to the iris. This study evaluates the reproducibility of optical sulcus measurements in the context of sizing a novel accommodative intraocular lens (IOL). Preoperative OCT scans of patients scheduled for cataract surgery were analyzed regarding the dimensions of the ciliary sulcus using a custom scan method with a clinically available anterior segment optical coherence tomographer. Measurements were compared between two different readers, and various derived parameters were compared. The measurements by both readers were highly correlated (R

Tópico:

Optical Coherence Tomography Applications

Citaciones:

Citations: 0
0

Citaciones por año:

No hay datos de citaciones disponibles

Altmétricas:

Paperbuzz Score: 0
0

Información de la Fuente:

SCImago Journal & Country Rank
FuenteSensors
Cuartil año de publicaciónNo disponible
Volumen24
Issue21
Páginas6950 - 6950
pISSNNo disponible
ISSNNo disponible

Enlaces e Identificadores:

Artículo de revista