4 Rings Electrical Impedance Tomography System Based on Analog Front-End Afe4300

Acceso Cerrado

Abstract:

A proposed low-cost, portable, 32-channel (4 rings of 8-channel) Electrical Impedance Tomography (EIT) system based on the AFE4300 analog front-end for body composition measurement. The switching of the 4 rings and the current injection and voltage measurement patterns are done with three Texas Instruments 74HC4067 multiplexers, which are managed by an ESP32 board. The proposed system has an average signal-to-noise ratio of 74.71 dB and a frame rate of 50 fps. The sensitivity tests to impedance and volume changes consisted of introducing 4 tubes of different diameters (2 steel and 2 polyvinyl chloride) into a tank with saline solution; then impedance distribution images were generated in 4 cross-sections of the tank, using the algorithms Gauss-Newton and Noser. Finally, the global impedance index (GI) is calculated to estimate the volume of each tube inside the tank. The results show that the proposed system is highly sensitive to impedance and volume changes, being a promising system for monitoring tissues, and fluids biological.

Tópico:

Electrical and Bioimpedance Tomography

Citaciones:

Citations: 0
0

Citaciones por año:

No hay datos de citaciones disponibles

Altmétricas:

Paperbuzz Score: 0
0

Información de la Fuente:

FuenteNo disponible
Cuartil año de publicaciónNo disponible
VolumenNo disponible
IssueNo disponible
PáginasNo disponible
pISSNNo disponible
ISSNNo disponible
Perfil OpenAlexNo disponible

Enlaces e Identificadores:

Artículo de revista