Effects of Variations in Operating Conditions on the Preci-sion and Accuracy of Standardless Elemental Analysis of Stainless Steel by SEM-EDS

Acceso Abierto

Abstract:

It is a customary practice to carry out standardless elemental microanalysis by energy dispersive spectroscopy (EDS) under the very same operational conditions as those used for scanning electron microscope (SEM) imaging. In this article, EDS experiments were carried out with stainless steel (SS), varying the EDS testing operating conditions. The results showed that, if X-ray spectra are acquired under operating conditions that are very dissimilar to those optimized for microanalysis, both the detectability limit of minor species (Si, Mo, and Mn) and the uncertainty in the concentration of major alloying elements (Cr and Ni) are noticeably impaired. It was observed that, by improving the signal-to-noise (S-to-N) ratio (i.e., by increasing the accelerating voltage, beam intensity, and total acquisition time, or when the working distance is optimized), the precision of the elemental concentration increases, but the accuracy is only marginally affected. For the major alloying elements, 25% of the measurements showed a percent discrepancy higher than three times the standard deviation, which is inconsistent with a normal statistical distribution.

Tópico:

Electron and X-Ray Spectroscopy Techniques

Citaciones:

Citations: 0
0

Citaciones por año:

No hay datos de citaciones disponibles

Altmétricas:

Paperbuzz Score: 0
0

Información de la Fuente:

SCImago Journal & Country Rank
FuenteIngeniería e Investigación
Cuartil año de publicaciónNo disponible
Volumen43
Issue1
Páginase94361 - e94361
pISSN0120-5609
ISSNNo disponible

Enlaces e Identificadores:

Artículo de revista