Logotipo ImpactU
Autor

Tribocorrosion Evaluation of Nb2O5, TiO2, and Nb2O5 + TiO2 Coatings for Medical Applications

Acceso Abierto

Abstract:

Materials used in biomedicine for purposes of long-time stay inside the body presents diverse sort of problems like cytotoxicity, wear, biocompatibility, and ion liberation along time. This paper presents the characterization of corrosion-wear combined phenomena on Nb2O5, TiO2, and Nb2O5 + TiO2 coatings with future applications as biomaterials. After the films’ production process using magnetron sputtering technique, they were characterized through an optic, scanning electron, and atomic force microscopy to evaluate their morphology, structure, and surface damage suffered by the synergy between wear and corrosion phenomena. The life in service of the implant was evaluated in terms of the coating behavior against inside body conditions like charge, wear, and electrochemical degradation. This test was made with electrochemical measurements in simulated biological fluid combined with the wear characterization implementing a potentiostat and a tribometer in a linear wear configuration with a bone pin. As a result, the different electrochemical responses of the films were evidenced by polarization curves and equivalent circuits of the systems. The coefficient of friction and surface degradation were also obtained and evaluated. Comparing the properties of the systems, we conclude that TiO2 coatings have better behavior in terms of the wear-corrosion synergy phenomena while the systems with Nb present pitting corrosion.

Tópico:

Titanium Alloys Microstructure and Properties

Citaciones:

Citations: 9
9

Citaciones por año:

Altmétricas:

Paperbuzz Score: 0
0

Información de la Fuente:

SCImago Journal & Country Rank
FuenteLubricants
Cuartil año de publicaciónNo disponible
Volumen9
Issue5
Páginas49 - 49
pISSNNo disponible
ISSNNo disponible

Enlaces e Identificadores:

Artículo de revista