Atomic Emission Spectrometry of Microsamples in a Glow Discharge

Acceso Cerrado

Abstract:

Abstract A glow discharge system for atomic emission spectrometry is evaluated for physical and analytical characteristics. Lead (as the nitrate) is the model element used for these studies. The sputtering characteristics of several electrode designs are studied with the best system tried resulting in total sample vaporization in less than 1 s. Discharge currents up to 200 mA at a discharge voltage of 1200 V and a pressure of 2 torr are found to give the best results. The detection limit for lead is 0.9 ng and the linear dynamic range of response is greater than 2 orders of magnitude. Sodium chloride at a 100-fold excess amount causes a 50% decrease in the signal level.

Tópico:

Analytical chemistry methods development

Citaciones:

Citations: 4
4

Citaciones por año:

No hay datos de citaciones disponibles

Altmétricas:

Paperbuzz Score: 0
0

Información de la Fuente:

SCImago Journal & Country Rank
FuenteSpectroscopy Letters
Cuartil año de publicaciónNo disponible
Volumen25
Issue8
Páginas1257 - 1265
pISSNNo disponible
ISSN1532-2289

Enlaces e Identificadores:

Artículo de revista