Abstract:

Use of occluder devices for a patent ductus arteriosus (PDA) closure is restricted to small diameter PDAs and involves high device costs. The objective of this study was to develop a new nitinol implant for the closure of a PDA and to evaluate its safety and efficacy in an animal model. The design consists of a device that promotes thrombus formation in the PDA. The device has a double-cone shape with their vertices joined. The cones are made of a nitinol wire helix with dacron fibers attached. The proximal end of the helix can be screwed to a conventional catheter release wire. In vitro testing was performed to verify the effectiveness of the implantation of the device and its releasing system; all trials (n = 30) were successful, showing accurate placement and release of the device. Complete and successful implantation of the device was achieved in all in vivo experiments (n = 5). There was one case of embolization due to premature detachment; a second device was successfully implanted. Histological evaluation after 42.3 +/- 3.1 days demonstrated complete PDA occlusion. The retrieved PDA showed total closure of the defect, endothelization of the PDA outlets and proper lodging of the device.

Tópico:

Cardiovascular Conditions and Treatments

Citaciones:

Citations: 3
3

Citaciones por año:

Altmétricas:

Paperbuzz Score: 0
0

Información de la Fuente:

SCImago Journal & Country Rank
FuenteASAIO Journal
Cuartil año de publicaciónNo disponible
Volumen55
Issue4
Páginas309 - 313
pISSNNo disponible
ISSN1538-943X

Enlaces e Identificadores:

Artículo de revista